Title of article
Robust registration for computer-integrated orthopedic surgery: Laboratory validation and clinical experience
Author/Authors
B. Ma، نويسنده , , R. E. Ellis، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2003
Pages
14
From page
237
To page
250
Abstract
In order to provide navigational guidance during computer-integrated orthopedic surgery, the anatomy of the patient must first be registered to a medical image or model. A common registration approach is to digitize points from the surface of a bone and then find the rigid transformation that best matches the points to the model by constrained optimization. Many optimization criteria, including a least-squares objective function, perform poorly if the data include spurious data points (outliers). This paper describes a statistically robust, surface-based registration algorithm that we have developed for orthopedic surgery. To find an initial estimate, the user digitizes points from predefined regions of bone that are large enough to reliably locate even in the absence of anatomic landmarks. Outliers are automatically detected and managed by integrating a statistically robust M-estimator with the iterative-closest-point algorithm. Our in vitro validation method simulated the registration process by drawing registration data points from several sets of densely digitized surface points. The method has been used clinically in computer-integrated surgery for high tibial osteotomy, distal radius osteotomy, and excision of osteoid osteoma.
Keywords
Computer-integrated orthopedic surgery , validation , Clinical experience
Journal title
Medical Image Analysis
Serial Year
2003
Journal title
Medical Image Analysis
Record number
449790
Link To Document