Automatic registration of portal images and volumetric CT for patient positioning in radiation therapy
Author/Authors
Ali Khamene، نويسنده , , Peter Bloch، نويسنده , , Wolfgang Wein، نويسنده , , Michelle Svatos، نويسنده , , Frank Sauer، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2006
Pages
17
From page
96
To page
112
Abstract
The efficacy of radiation therapy treatment depends on the patient setup accuracy at each daily fraction. A significant problem is reproducing the patient position during treatment planning for every fraction of the treatment process. We propose and evaluate an intensity based automatic registration method using multiple portal images and the pre-treatment CT volume. We perform both geometric and radiometric calibrations to generate high quality digitally reconstructed radiographs (DRRs) that can be compared against portal images acquired right before treatment dose delivery. We use a graphics processing unit (GPU) to generate the DRRs in order to gain computational efficiency. We also perform a comparative study on various similarity measures and optimization procedures. Simple similarity measure such as local normalized correlation (LNC) performs best as long as the radiometric calibration is carefully done. Using the proposed method, we achieved better than 1 mm average error in repositioning accuracy for a series of phantom studies using two open field (i.e., 41 cm2) portal images with 90° vergence angle.
Keywords
Patient positioning , 2D–3D registration , radiation therapy , DRR generation