Title of article
Accuracy evaluation of direct navigation with an isocentric 3D rotational X-ray system
Author/Authors
Everine B. van de Kraats، نويسنده , , Theo van Walsum، نويسنده , , Lance Kendrick، نويسنده , , Niels J. Noordhoek، نويسنده , , Wiro J. Niessen، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2006
Pages
12
From page
113
To page
124
Abstract
Minimally invasive interventions are often performed under fluoroscopic guidance. Drawbacks of fluoroscopic guidance are the fact that the presented images are 2D projections and that both the patient and the clinician are exposed to radiation. Image-guided navigation using pre-interventionally acquired 3D MR or CT data is an alternative. However, this often requires invasive anatomical landmark-based, marker-based or surface-based image-to-patient registration. In this paper, a coupling between an image-guided navigation system and an intraoperative C-arm X-ray device with 3D imaging capabilities (3D rotational X-ray (3DRX) system) that enables direct navigation without invasive image-to-patient registration on 3DRX volumes, is described and evaluated. The coupling is established in a one-time preoperative calibration procedure. The individual steps in the registration procedure are explained and evaluated. The acquired navigation accuracy using this coupling is approximately one millimeter.
Keywords
Direct navigation , calibration , accuracy , 3D rotational X-ray imaging , minimally invasive surgery
Journal title
Medical Image Analysis
Serial Year
2006
Journal title
Medical Image Analysis
Record number
449904