Loads on an internal spinal fixation device during walking
Author/Authors
Antonius Rohlmann، نويسنده , , Georg Bergmann، نويسنده , , Friedmar Graichen، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1997
Pages
7
From page
41
To page
47
Abstract
Only little knowledge exists concerning the loads on internal spinal fixation devices during walking. In this study, forces and moments were measured in two patients using telemeterized spinal fixators. Although implant loads differed strongly before and after anterior fusion as well as between the two patients, some results were consistent. In every test series, implant loads were higher in walking than in lying, sitting or standing. Walking speed had little influence on implant loads. Staircase walking put slightly higher loads on the implants than normal level walking. Normal use of two crutches reduced implant loads only slightly, whereas a wheeled invalid walker reduced them by about 25%.
Keywords
Spine: Spinal fixation device , Walking , Load measurement , Crutches