Title of article :
A method to determine the 3-D stiffness of fracture fixation devices and its application to predict inter-fragmentary movement
Author/Authors :
Georg N. Duda، نويسنده , , Helmut Kirchner، نويسنده , , Hans-Joachim Wilke، نويسنده , , Lutz Claes، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Pages :
6
From page :
247
To page :
252
Abstract :
Inter-fragmentary movement considerably influences the fracture healing process. Large shear movement delays while moderate axial movement stimulates the healing process. To be able to control the mechanical situation at a fracture site and to achieve optimal bony healing it is essential to understand the relationship between inter-fragmentary movement, bony loading and fixation stiffness. A 6×6 stiffness matrix is introduced which completely describes the linear relationship between the 6 inter-fragmentary movements and the resulting bony loading (3 forces and 3 moments). Further, it is illustrated that even in relatively stiff external fixateur constructs simple axial loading of the bony fragments leads to complex inter-fragmentary movement. When the 3-D stiffness description is multiplied by the load state in sheep tibiae, movements similar to those measured in vivo are calculated. The relationship between axial compression and medio-lateral or dorso-ventral shear varies depending on the mounting plane of the external fixateur. The authors conclude that a single value is not sufficient to describe the mechanical relationship between inter-fragmentary movement and bony loading. Only a complete description of fixation stiffness allows prediction of inter-fragmentary movement and differentiation between various configurations of fixation devices and their potential for mechanically promoting bony healing.
Keywords :
External ?xation , Fracture Healing , Inter-fragmentary movement , Fixation sti¤ness
Journal title :
Journal of Biomechanics
Serial Year :
1998
Journal title :
Journal of Biomechanics
Record number :
450667
Link To Document :
بازگشت