Title of article
Hip endoprosthesis for in vivo measurement of joint force and temperature
Author/Authors
Friedmar Graichen، نويسنده , , Georg Bergmann، نويسنده , , Antonius Rohlmann، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1999
Pages
5
From page
1113
To page
1117
Abstract
Friction between the prosthetic head and acetabular cup increases the temperature in hip implants during activities like walking. A hip endoprosthesis was instrumented with sensors to measure the joint contact forces and the temperature distribution along the entire length of the titanium implant. Sensors and two inductively powered telemetry units are placed inside the hip implant and hermetically sealed against body fluids. Each telemetry unit contains an integrated 8-channel telemetry chip and a radio frequency transmitter. Force, temperature and power supply data are transmitted at different frequencies by two antennas to an external twin receiver. The inductive power supply is controlled by a personal computer. Force and temperature are monitored in real time and all data are stored on a video tape together with the patientʹs images. This paper describes the design and accuracy of the instrumented implant and the principal function of the external system components.
Keywords
telemetry , Hip joint force , In vivo measurement , Hip joint temperature , biomechanics , Instrumented hip prosthesis
Journal title
Journal of Biomechanics
Serial Year
1999
Journal title
Journal of Biomechanics
Record number
450922
Link To Document