Title of article :
A continuous pure moment loading apparatus for biomechanical testing of multi-segment spine specimens
Author/Authors :
John T. Lysack، نويسنده , , James P. Dickey، نويسنده , , Genevieve A. Dumas، نويسنده , , C. David Yen، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Abstract :
An apparatus is described that enables the application of continuous pure moment loads to multi-segment spine specimens. This loading apparatus allows continuous cycling of the spine between specified flexion and extension (or right and left lateral bending) maximum load endpoints. Using a six-degree-of-freedom load cell and three-dimensional optoelectronic stereophotogrammetry, characteristic displacement versus load hysteresis curves can be generated and analyzed for different spinal constructs of interest. Unlike quasi-static loading, the use of continuous loading permits the analysis of the spineʹs behaviour within the neutral zone. This information is of particular clinical significance given that the instability of a spinal segment is related to its flexibility within the neutral zone. Representative curves for the porcine lumbar spine in flexion–extension and lateral bending are presented to illustrate the capabilities of this system.
Keywords :
Spine , Loading apparatus , Flexibility testing , Continuous loads , Pure moments
Journal title :
Journal of Biomechanics
Journal title :
Journal of Biomechanics