Title of article :
A three-dimensional computational analysis of fluid–structure interaction in the aortic valve
Author/Authors :
J. de Hart، نويسنده , , G. W. M. Peters، نويسنده , , P. J. G. Schreurs ، نويسنده , , F. P. T. Baaijens، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Abstract :
Numerical analysis of the aortic valve has mainly been focused on the closing behaviour during the diastolic phase rather than the kinematic opening and closing behaviour during the systolic phase of the cardiac cycle. Moreover, the fluid–structure interaction in the aortic valve system is most frequently ignored in numerical modelling. The effect of this interaction on the valveʹs behaviour during systolic functioning is investigated. The large differences in material properties of fluid and structure and the finite motion of the leaflets complicate blood–valve interaction modelling. This has impeded numerical analyses of valves operating under physiological conditions. A numerical method, known as the Lagrange multiplier based fictitious domain method, is used to describe the large leaflet motion within the computational fluid domain. This method is applied to a three-dimensional finite element model of a stented aortic valve. The model provides both the mechanical behaviour of the valve and the blood flow through it. Results show that during systole the leaflets of the stented valve appear to be moving with the fluid in an essentially kinematical process governed by the fluid motion.
Keywords :
aortic valve , Finite element method , Fictitious domain method , Three-dimensional analysis , Fluid–structure interaction
Journal title :
Journal of Biomechanics
Journal title :
Journal of Biomechanics