• Title of article

    A cryo-jaw designed for in vitro tensile testing of the healing Achilles tendons in rats

  • Author/Authors

    Peter Wieloch، نويسنده , , Gerhard Buchmann، نويسنده , , Wolfgang Roth، نويسنده , , Markus Rickert، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    4
  • From page
    1719
  • To page
    1722
  • Abstract
    We present a simple but very secure non-compressive clamping device for in vitro tensile testing of the Achilles tendon in rats. Biomechanical studies were performed on 178 native and injured specimens without any registration of slippage out of the clamp. The average tensile strength of the native rat Achilles tendon was determined to be 48 N (±11 N) and its displacement to failure averaged 0.8 mm (±0.2 mm). The mean values for tensile strength within the injured tendons increased with time and were above the native tendon values from week 2 and later. Correspondingly, the displacement to failure decreased with time.
  • Keywords
    Tensile testing , Rat Achilles tendon , Tendon healing , Cryo-jaw
  • Journal title
    Journal of Biomechanics
  • Serial Year
    2004
  • Journal title
    Journal of Biomechanics
  • Record number