Title of article :
Measurements of constraint of total knee replacement
Author/Authors :
H. Haider، نويسنده , , P. S. Walker، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Pages :
8
From page :
341
To page :
348
Abstract :
Measurement of the constraint of total knee components in a test machine provides an objective method of describing the laxity and stability characteristics of the implant itself, independent of the knee joint into which it would be implanted. A special fixture was designed and fitted to a Bionix multi-channel loading machine. The test consisted of applying a compressive load, applying a cyclic AP force or internal-external torque, and measuring all of the displacements and rotations. Three different commonly-used TKRʹs showed widely different constraint characteristics. In the cyclic AP test, along with the cyclic AP displacement, displacements and rotations occurred in the other directions. This indicated that all degrees of freedom should be free to move, otherwise anomalous results would be obtained. The paper concludes with recommendations for standardized constraint tests.
Keywords :
Total knee replacement (TKR) , TKR constraint , TKR laxity , TKR stability , TKR kinematics , TKR testing
Journal title :
Journal of Biomechanics
Serial Year :
2005
Journal title :
Journal of Biomechanics
Record number :
451951
Link To Document :
بازگشت