• Title of article

    In vitro fatigue behaviour of a cemented acetabular reconstruction

  • Author/Authors

    Phillip Heaton-Adegbile، نويسنده , , Nikolaus P. Zant، نويسنده , , Jie Tong، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    5
  • From page
    2882
  • To page
    2886
  • Abstract
    In this study, a hemi-pelvis of composite sawbone was implanted with a Charnley cup using a conventional bone cement and the acetabular replacement was tested under constant amplitude cyclic loads, simulating the maximum hip contact force during normal walking. The damage development in the reconstruction was detected and monitored using CT scanning at regular test intervals, verified by microscopy post testing. Three identical experimental results showed that extensive debonding at the bone–cement interface occurred around the dome region after 20 million cycles.
  • Keywords
    Cemented fixation , CT scans , Implant loosening , Fatigue damage
  • Journal title
    Journal of Biomechanics
  • Serial Year
    2006
  • Journal title
    Journal of Biomechanics
  • Record number

    452358