Title of article :
Effect of bearing geometry and structure support on transient elastohydrodynamic lubrication of metal-on-metal hip implants
Author/Authors :
Feng Liu، نويسنده , , Zhongming Jin، نويسنده , , Paul Roberts، نويسنده , , Peter Grigoris، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Pages :
10
From page :
1340
To page :
1349
Abstract :
An effective lubrication can significantly reduce wear of metal-on-metal artificial hip joints. The improvement of the lubrication can be achieved through the optimisation of the bearing geometry in terms of a small clearance and/or the structural support such as a polyethylene backing underneath a metallic bearing in a sandwich acetabular cup form. The separate effects of these two factors on fluid film lubrication of 28 mm diameter metal-on-metal total hip joints under walking conditions were numerically investigated in this paper. The results show that a larger lubricant film due to the polyethylene backing can be significantly enhanced by the transient squeeze-film action, particularly during the stance phase, and a similar lubricant film can be developed for both the monolithic cup relying on the smaller clearance and the sandwich cup benefiting from the polyethylene backing. Both cup systems can function in a wide range of lubrication regimes, covering both mixed and fluid film, under the current design and manufacture conditions.
Keywords :
Transient elastohydrodynamic lubrication , Metal-on-metal hip implants , UHMWPE backing
Journal title :
Journal of Biomechanics
Serial Year :
2007
Journal title :
Journal of Biomechanics
Record number :
452546
Link To Document :
بازگشت