Title of article :
Primary rotational stability of cylindrical and conical revision hip stems as a function of femoral bone defects: An in vitro comparison
Author/Authors :
Eike Jakubowitz، نويسنده , , Rudi G. Bitsch، نويسنده , , Christian Heisel، نويسنده , , Christoph Lee، نويسنده , , Jan P. Kretzer، نويسنده , , Marc N. Thomsen، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Abstract :
Bone stock losses in cementless femoral stem revisions compromise a stable fixation. The surgeon has to rely on his wealth of experience in deciding which stem shape to use. The aim of our study was to compare the primary rotational stability of cylindrical and conical revision hip stems subjected to femoral defects. Four current prostheses (two cylindrical, two conical) were implanted into four synthetic femora. Micro-motion was measured under torque application and femoral neck osteotomy and segmental AAOS Type I and III defects were simulated. The relative movements of all prostheses were significantly influenced by the extent of bone loss (p<0.01). Major differences were seen in fixation behavior (p<0.01). The main fixation area of conical stems is within the distal femoral isthmus, whereas cylindrical implants are dependent on proximal bone stock. In our study, cylindrical stems are advantageous for minor defects because they provide a proximal fixation. In cases of extensive substance loss, the conical implants showed lesser relative movements. These findings should be taken into account for clinical decisions.
Keywords :
Primary stabilityRevision hip stemStem shapeFemoral defectBridging
Journal title :