Title of article :
Evaluation of a new algorithm to determine the hip joint center
Author/Authors :
Robert A. Siston، نويسنده , , Scott L. Delp، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Pages :
6
From page :
125
To page :
130
Abstract :
Accurately locating the hip joint center is a challenging and important step in many biomechanical investigations. The purpose of this study was to test the accuracy and robustness of a “pivoting” algorithm used to locate the hip center. We tested the performance of this algorithm with data acquired by manipulating a ball and socket model of the hip through several motion patterns. The smallest mean errors of 2.2±0.2 mm occurred with a circumduction motion pattern, while the largest errors of 4.2±1.3 mm occurred with single-plane motion (e.g., flexion/extension). Introducing random noise with an amplitude of 30 mm increased the errors by only 1.3±0.5 mm with a circumduction motion pattern. The pivoting algorithm performs well in the laboratory, and further work is warranted to evaluate its performance in a clinical setting.
Keywords :
Hipcenter , Motion analysis , Kinematics , computer-assisted surgery
Journal title :
Journal of Biomechanics
Serial Year :
2006
Journal title :
Journal of Biomechanics
Record number :
453368