The Hydro ThermAblator
Author/Authors
Stephen L. Corson، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2007
Pages
9
From page
979
To page
987
Abstract
The Hydro ThermAblator system is a unique approach to endometrial ablation, employing constant direct visual control via hysteroscopy combined with a ‘no-touch’ gravity flow of heated saline that treats the entire endometrial surface regardless of its topography. The automated controller helps to ensure reproducible results without the necessity for high operator skill. The system employs a safety feature that monitors the fluid level constantly within a closed system, so that loss of fluid through unsuspected uterine perforation occurring during dilatation can be diagnosed even though it may not be visually apparent to the operator. Results as judged by patient satisfaction, amenorrhoea rates and/or return to normal menstrual flow are excellent.