  • Abstract
    The aim of the study was to compare the potassium efflux measurements (flame photometry (FP), specific electrode (SE) and atomic absorption photometry (AAP)) using a model of erythrocytes exposed to an oxidative stress in various conditions of osmolarity. Human erythrocytes were incubated in 3 different values of osmolarity and in the presence of 50 mM AAPH, potassium efflux was measured by FP, SE and AAP at t = 0 and every 30 min for 2 h. These methods were similar for the measurement of global potassium efflux. However, SE detected important amounts of potassium at the beginning of the experiment or in absence of AAPH in comparison with AAP and FP. It is noteworthy that these different methods of measurements were not altered by the osmolarity. FP and AAP make it possible to study the potassium efflux during oxidative stress while SE should be used only for global measurements.
  • Keywords
    potassium , Erythrocytes , AAPH