Title of article :
Intraoperative validation of mitral inflow determination by transesophageal echocardiography: Comparison of single-plane, biplane and thermodilution techniques
Author/Authors :
Pu Min، نويسنده , , Brian P. Griffin، نويسنده , , Pieter M. Vandervoort، نويسنده , , Dominic Y. Leung، نويسنده , , Delos M. Cosgrove III، نويسنده , , James D. Thomas، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1995
Abstract :
Objectives.
This study investigated the accuracy of mitral inflow quantification using biplane transesophageal echocardiography.
Background.
Mitral stroke volume can be reliably quantified by transthoracic Doppler echocardiography, but previous studies involving monoplane transesophageal echocardiography have yielded mixed results.
Methods.
Thirty patients without mitral regurgitation were prospectively examined immediately before cardiovascular surgery. Mitral annulus diameter was measured in the transverse (d1) and longitudinal views (d2) by biplane transesophageal echocardiography. Assuming an elliptic shape, the annular are was calculated as πd1d2/4; are was also calculated from single-plane dat assuming circular annular shape as πd24. The time—velocity integral of mitral annular Doppler velocity was then multiplied by annular are to yield stroke volume. These dat were compared with simultaneous thermodilution measurements by linear regression.
Results.
Good correlations were observed between thermodilution (x) and Doppler (y) measurements of stroke volume (SV) (r = 0.86, p < 0.01, ΔSV [y−x]= 2.64 ± 9.86 ml for single four-chamber view; r = 0.77, p < 0.01, ΔSV = 1.82 ± 12.59 ml for two-chamber view; r = 0.94, p < 0.001, ΔSV = 1.78 ± 5.90 ml for biplane measurements) with similar dat for cardiac output (r = 0.82, r = 0.74 and r = 0.92, respectively). The biplane measurements were most accurate and had less variability in individual patients (p < 0.05). This finding was supported by numerical model that demonstrated (for an ellipse of eccentricity 1.5: 1) that even maximal misalignment of biplane diameters yielded only 8% are overestimation, whereas single-plane calculations assuming circular shape produced variation in are of 225%.
Conclusions.
This study validates the accuracy of measurements of mitral inflow using biplane transesophageal echocardiography with potential application for quantification of valvular regurgitation in the operating room. The results are further generalizable, indicating that orthogonal biplane measurements are both necessary and sufficient to ensure accuracy in are calculation for any elliptic structure.
Journal title :
JACC (Journal of the American College of Cardiology)