Title of article :
Catheter Closure of Moderate- to Large-Sized Patent Ductus Arteriosus Using the New Amplatzer Duct Occluder: Immediate and Short-Term Results
Author/Authors :
Jozef Masur MD PhD، نويسنده , , Kevin P. Walsh MD MRCP، نويسنده , , Basil Thanopoulous MD، نويسنده , , Chen Chan MD، نويسنده , , John Bass MD، نويسنده , , Yousef Goussous MD FACC، نويسنده , , Paul Gavor MD، نويسنده , , Ziyad M. Hijazi MD MPH، نويسنده , , FACC، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Pages :
5
From page :
878
To page :
882
Abstract :
Objectives. The aim of this study was to assess the immediate and short-term results of anterograde catheter closure of moderate- to large-sized patent ductus arteriosus (PDA) using the new self-expandable, respositionable Amplatzer duct occluder (ADO) device. Background. Transcatheter closure of PD using devices or coils is technically challenging and may be accompanied by 38% incidence of residual shunts. Methods. Twenty-four patients (6 male, 18 female) underwent attempted transcatheter closure of PD using the ADO at median age of 3.8 years (range 0.4 to 48) and median weight of 15.5 kg (range 6 to 70). The mean PD diameter at its narrowest segment was 3.7 ± 1.5 mm. 6F long sheath was used for delivery of the ADO. Follow-up evaluation was performed with color flow mapping of the main pulmonary artery within 24 h and at 1 and 3 months after closure. Results. Twenty three of the 24 patients had successful device placement. Angiography showed that 7 patients had complete immediate closure, 14 had trace shunt (foaming through the device with no jet), and 2 had small residual shunt (with jet). Within 24 h, color Doppler revealed complete closure in all patients. The unsuccessful attempt was during an initial trial with prototype that has been modified. The median fluoroscopy time was 13.5 min (range 6.3 to 47). All patients were discharged home the next day. There were no complications. Of the 23 patients, 21 completed the 1-month follow-up, all (95% confidence interval [CI] 86% to 100%) with complete closure, and 18 of 23 patients completed the 3-month follow-up, also all (95% CI 83% to 100%) with complete closure. Conclusions. Anterograde transcatheter closure using the new ADO is an effective therapy for patients with PD diameter up to 6 mm. Further clinical trials are underway.
Keywords :
PDA , Confidence interval , CI , Qp/Qs , patent ductus arteriosus , pulmonary/systemic flow ratio , ADO , Amplatzer Duct Occluder
Journal title :
JACC (Journal of the American College of Cardiology)
Serial Year :
1998
Journal title :
JACC (Journal of the American College of Cardiology)
Record number :
480625
Link To Document :
بازگشت