Title of article
The influence of ventricular fibrillation duration on defibrillation efficacy using biphasic waveforms in humans
Author/Authors
Stephan Windecker*، نويسنده , , Raymond E Ideker، نويسنده , , Vance J Plumb، نويسنده , , G.Neal Kay، نويسنده , , Gregory P. Walcott، نويسنده , , Andrew E Epstein، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1999
Pages
6
From page
33
To page
38
Abstract
Objectives. The purpose of this study was to prospectively investigate the influence of ventricular fibrillation (VF) durations of 5, 10 and 20 s on the defibrillation threshold (DFT) during implantable cardioverter–defibrillator (ICD) implantation.
Background. Although the DFT using monophasic waveforms has been shown to increase with VF duration in humans, the effect of VF duration on defibrillation efficacy using biphasic waveforms in humans is not known.
Methods. Thirty patients undergoing primary ICD implantation or pulse generator replacement were randomly assigned to have the DFT determined using biphasic shocks at two durations of VF each (5 and 10 s, 10 and 20 s or 5 and 20 s).
Results. There was no statistically significant difference in the mean DFT comparing VF durations of 5 s (9.5 ± 6.0 J) and 10 s (10.8 ± 7.0 J) (p = 0.4). The mean DFT significantly increased from 10.9 ± 6.1 J at 10 s of VF to 12.6 ± 5.6 J (p = 0.03) at 20 s of VF, and from 7.0 ± 3.5 J at 5 s of VF to 10.5 ± 6.3 J (p = 0.04) at 20 s of VF. An increase in the DFT was observed in 14 patients as VF duration increased. There were no clinical characteristics that differentiated patients with and without an increase in the DFT.
Conclusions. Defibrillation efficacy decreases with increasing VF duration using biphasic waveforms in humans. Ventricular fibrillation durations greater than 10 s may negatively affect the effectiveness of ICD therapy.
Keywords
DFT , ICD , Vf , ventricular fibrillation , defibrillation threshold , implantable cardioverter–defibrillator
Journal title
JACC (Journal of the American College of Cardiology)
Serial Year
1999
Journal title
JACC (Journal of the American College of Cardiology)
Record number
480987
Link To Document