Title of article :
Assay of neonatal TSH from blood spots using the Architect i2000, an immunoassay analyzer
Author/Authors :
Joyce LePage، نويسنده , , Sherry Peters، نويسنده , , Denis C. Lehotay، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Pages :
3
From page :
415
To page :
417
Abstract :
Objectives: To assess the Architect i2000 for measuring TSH from neonatal blood spots. Methods: Blood collected on filter paper was eluted and assayed on the Architect. Results: Recovery of TSH was close to 100%. Within-run and between-run precisions were 4.2–8.2% and 14.5–18.1%. Correlation with the Wallac Delphia method was excellent (r = 0.96). Linearity, sensitivity, and slope were 0–400 mU/l, 0.15 mU/l, and 0.86, respectively. The results obtained by this method on the Architect allowed us to differentiate normals from affected patients with the same reliability as the previous method. Conclusions: Measurement of TSH on the Architect is an excellent method for detecting congenital hypothyroidism.
Keywords :
TSH , Architect i2000 , Radioimmunoassays
Journal title :
Clinical Biochemistry
Serial Year :
2004
Journal title :
Clinical Biochemistry
Record number :
482534
Link To Document :
بازگشت