Title of article :
Two-center clinical evaluation of a new automated fluorometric immunoassay for the quantitative analysis of total βeta-human chorionic gonadotropin
Author/Authors :
Hassan M. E. Azzazy، نويسنده , , Liana Frances Romero، نويسنده , , Bruce Lee Hall، نويسنده , , Doris Ruttle، نويسنده , , Robert H. Christenson، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
6
From page :
523
To page :
528
Abstract :
Objectives This study evaluated the quantitative total βhCG assay on the Stratus® CS point-of-care instrument at two medical centers. Design and methods Analytical sensitivity, linearity, within-run and total imprecision, interferences, dilution recovery, method comparison (Dimension® RxL), comparison of matched heparinized whole blood and plasma samples, and determination of the normal reference interval were studied. Results Analytical sensitivity was <0.5 IU/L. The assayʹs linear range was 0 to 1250 IU/L; the clinical reportable range was up to 50,000 IU/L. Within-run imprecision (CV) at both low (<20 IU/L) and elevated (760 IU/L) βhCG concentrations were <4%. Total imprecision for three QC levels and two pools were <4%. Method comparison showed Stratus CS βhCG = 0.98 ± 0.01* Dimension RxL hCG −0.11 ± 2.69 (n = 136; r = 0.996; Sy/x = 27.7). Matched heparinized whole blood/plasma sample-comparison showed: whole BLOOD = 1.05*Plasma + 0.37 ± 1.29 (n = 41; r = 1.000; Sy/x = 7.57). Mean dilution recovery was 99% (range: 95% to 103%). None of the 52 drugs tested, lipemia, icterus, hemolysis, LH, FSH, TSH, hGH or prolactin represented a significant interference with the assay. Reference intervals were <0.5 IU/L for males (n = 123) and <3.0 IU/L for nonpregnant females (n = 120). Conclusions The Stratus® CS βhCG test offers the advantage of quantitative measurement of total βhCG in whole blood at the point of care and is suitable for clinical use.
Keywords :
quantitative , Total beta human chorionic gonadotropin , Fluorometric immunoassay , Pregnancy monitoring , Stratus CS
Journal title :
Clinical Biochemistry
Serial Year :
2003
Journal title :
Clinical Biochemistry
Record number :
484640