Title of article :
Comparative study between the Light Cycler® and the PCR-restriction fragment length polymorphism in detecting factor V Leiden and factor II 20210G>A mutations
Author/Authors :
Layla Mammo، نويسنده , , Atia Sheereen، نويسنده , , Tanya Saour، نويسنده , , Jalal Saour، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Abstract :
Objectives
To test reproducibility, speed and cost of testing for factor V Leiden and FII 20210G>A in our practice.
Design and methods
We compared conformity, reproducibility, speed and cost using the Light Cycler (LC®) and PCR-RFLP.
Results
There was 100% conformity and reproducibility. LC® was faster but 23% more expensive per sample. When equipment depreciation and patient expenses are added, LC® testing becomes cheaper.
Conclusion
In our practice, LC® provides fast, reproducible and cost-effective results.
Keywords :
polymerase chain reaction (PCR) , Factor V Leiden (FVL) , Factor II 20210G>A (FII 20210G>A) , Roche Light Cycler (LC®) , Restriction FragmentLength Polymorphism (RFLP)
Journal title :
Clinical Biochemistry
Journal title :
Clinical Biochemistry