Title of article :
A new turbidimetric standard to improve the quality assurance of the erythrocyte sedimentation rate measurement
Author/Authors :
Elisa Piva، نويسنده , , Rachele Pajola، نويسنده , , Valeria Temporin، نويسنده , , Mario Plebani، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Abstract :
Objectives:
The erythrocyte sedimentation rate (ESR) determined using manual or semi-automated methods has usually been judged a simple procedure that could be performed without any form of Quality Control. According to this point of view, the ESR would seem a semi-quantitative test instead of a real hematological test with any clinical importance. The new millennium has consolidated ESR automation and a QC plan cannot be postponed.
Design and methods:
Using Test1, an automated ESR analyzer, a new latex control material was evaluated and compared with fresh whole blood for quality control purposes.
Results:
The new latex control showed satisfactory reproducibility, precision, and “commutability” in comparison to quality control procedures that use fresh whole blood samples.
Conclusions:
The new standard represents another advance in ESR testing improving the accuracy of automated TEST1 measurements.
Keywords :
quality control , quality assurance , erythrocyte sedimentation rate , CLSI , TEST1 analyzer
Journal title :
Clinical Biochemistry
Journal title :
Clinical Biochemistry