Title of article :
Measurement of Elecsys NT-proBNP in serum, K2 EDTA and heparin plasma
Author/Authors :
Giuseppe Lippi، نويسنده , , Gian Luca Salvagno، نويسنده , , Martina Montagnana، نويسنده , , Gian Cesare Guidi، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Pages :
2
From page :
747
To page :
748
Abstract :
Objective: There is controversial evidence on a matrix influence on the measurement of NT-proBNP. Design and methods: We compared results of Elecsys NT-proBNP measurement on serum, K2 EDTA plasma and lithium heparin plasma. Results: Samples collected in K2 EDTA showed a marginally significant underestimation when compared to serum and heparin, whereas no significant difference was observed between serum and heparin plasma. Conclusions: Serum, heparin and K2 EDTA plasma may be suitable for NT-proBNP measurement.
Keywords :
serum , heparin , EDTA , Natriuretic peptides