Title of article :
Head-to-head comparison of an automated immunometric and an automated HPLC method for the quantification of HbA1c
Author/Authors :
Michael Vogeser، نويسنده , , Klaus G. Parhofer، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Pages :
3
From page :
1410
To page :
1412
Abstract :
Objectives To compare a fully automated immunoassay and a fully automated HPLC method for the quantification of HbA1c. Design and methods For an inter-method comparison, 200 consecutive clinical samples were analyzed. Imprecision was studied with pooled whole blood. Results Very close agreement of results (identical mean result (6.0%); r = 0.99), and very similar reproducibility (total CV approximately 1%) were observed. Conclusions Both assays offer very similar analytical performance, although the analytical techniques used are completely different.
Keywords :
Immunoassay , HPLC , Turbidimetry , glycated haemoglobin , HbA1C
Journal title :
Clinical Biochemistry
Serial Year :
2008
Journal title :
Clinical Biochemistry
Record number :
485355
Link To Document :
بازگشت