Title of article :
Three-dimensional primary stability of cementless femoral stems
Author/Authors :
D. W. Bühler، نويسنده , , Michael Berlemann، نويسنده , , K Lippuner، نويسنده , , P Jaeger، نويسنده , , L. P. Nolte، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1997
Abstract :
Objective. This study investigates by means of a new bone-prosthesis interface motion detector whether conceptual design differences of femoral stems are reflected in their primary stability pattern.
Design. An in vitro experiment using a biaxial materials testing machine in combination with three-dimensional motion measurement devices was performed.
Background. Primary stability of uncemented total hip replacements is considered to be a prerequisite for the quality of bony ongrowth to the femoral stem. Dynamic motion as a response to loading as well as total motion of the prosthesis have to be considered under quasi-physiological cyclic loading conditions.
Methods. Seven paired fresh cadaveric femora were used for the testing of two types of uncemented femoral stems with different anchoring concepts: CLS stem (Spotorno) and Cone Prosthesis (Wagner). Under sinusoidal cyclic loading mimicking in vivo hip joint forces a new measurement technique was applied allowing for the analysis of the three-dimensional interface motion.
Results. Considerable differences between the two prostheses could be detected both in their dynamic motion and total motion behaviour. Whereas the CLS stem, due to the wedge-shaped concept, provides smaller total motions, the longitudinal ribs of the Cone prostheses result in a substantially smaller dynamic motion.
Conclusions. The measuring technique provided reliable and accurate data illustrating the three-dimensional interface motion of uncemented femoral stems.
Keywords :
Primary stability , interface motion , Prosthesis design , in vitro , Mechanical testing , Total hip replacement
Journal title :
Clinical Biomechanics