• Title of article

    Rapid application fracture fixators – an evaluation of mechanical performance

  • Author/Authors

    Trevor N. Gardner، نويسنده , , H. Simpson، نويسنده , , J. Kenwright، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2001
  • Pages
    9
  • From page
    151
  • To page
    159
  • Abstract
    Objective. This study evaluates the mechanical performance of the Pinless and Centrafix fixators for rapid application to tibial fractures in a disaster or battlefield scenario. Design. Comparative study based on measurements made in the laboratory. Background. The Pinless and Centrafix fixators may be considered for rapid application to stabilise fractures in emergency conditions without the aid of electrical equipment such as power drills for bone screw insertion or image intensifiers to facilitate bone alignment. Methods. Stiffnesses, maximum service loads and fatigue strengths of the fixators were measured in the orientations of loading that correspond to walking and stretcher-bearing. These properties were compared with measurements on three conventional fixators, the AO, Shearer and Triax. Results. The Centrafix stiffnesses were 31 N/mm (axial), 1 N/°(torsional shear), 0.4 N/°(coronal plane bending), 4 N/°(sagittal plane bending) and 11 N/mm (transverse shear) and strengths were 95 N (axial) and 1.9 Nm (bending). Corresponding Pinless stiffnesses were 43 N/mm, 0.7 N/°, 0.3 N/°, 8 N/° and 50 N/mm, and strength was 55 N (axial). Conclusions. The stiffness and strength of both rapid application fixators in simulated walking was judged to be low, and additionally the stiffness and strength of the Centrafix in simulated stretcher-bearing was judged to be low.
  • Keywords
    fracture , External fixator , Centrafix , Stiffness , Strength , Pinless
  • Journal title
    Clinical Biomechanics
  • Serial Year
    2001
  • Journal title
    Clinical Biomechanics
  • Record number

    485903