Title of article :
The effect of metallic radial head arthroplasty on radiocapitellar joint contact area
Author/Authors :
Victor S. Liew، نويسنده , , Ian C. Cooper، نويسنده , , Louis M. Ferreira، نويسنده , , James A. Johnson، نويسنده , , Graham J. W. King، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Abstract :
Objective. This study determined the effect of metallic radial head arthroplasty on radiocapitellar joint contact area.
Design. The contact area of eight intact radiocapitellar joints was assessed during simulated compressive loading. Subsequently, the same experiments were repeated after reconstruction with three different radial head implant sizes.
Background. There has been an increasing use of metal radial head implants relative to silicone implants. However, the contact characteristics with the metal reconstruction have not been determined.
Methods. A 100 N compressive load was applied to the radiocapitellar joint at three different flexion angles for the native joint and three different sizes of a metal head implant. An impression material was employed to quantify joint contact area.
Results. Following metallic radial head arthroplasty, the contact area decreased by approximately two-thirds relative to the native radiocapitellar joint (P<0.001). Smaller implant sizes resulted in slightly larger contact areas (P<0.05). In addition, contact area decreased as joint flexion angle increased (P<0.04).
Conclusions. Selection of radial head implant size is likely not a significant factor with regard to contact area mechanics of the radiocapitellar articulation.
Keywords :
Metal radial head implant , Radial head fracture , Joint casting , Elbow joint contact mechanics
Journal title :
Clinical Biomechanics