• Title of article

    Axial compression force measurement acting across the strut graft in thoracolumbar instrumentation testing

  • Author/Authors

    Markus Schultheiss، نويسنده , , Erich Hartwig، نويسنده , , Lothar Kinzl، نويسنده , , Lutz Claes، نويسنده , , Hans-Joachim Wilke، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    6
  • From page
    631
  • To page
    636
  • Abstract
    Objective. Current recommendations for spinal implant testing do not consider the determination of axial compression forces of the overbridging implant on the strut graft. No direct data exist on the influence of load transfer through the strut graft and of the kind of instrumentation, especially in thoracolumbar corpectomy models. Design. Therefore in this biomechanical in vitro study a method for measurement of the axial compression force acting across the strut graft in different thoracolumbar instrumentations was developed. Methods. In this in vitro study, a corpectomy model was simulated and anterior, posterior and combined short fixation devices currently available were tested under pure moments to evaluate their biomechanical stabilizing characteristics. Range of motion, neutral zone and the axial compressive force acting on the strut graft were measured continuously in the three primary directions. Results. Without loads, the combined stabilization and followed by anterior instrumentation created a higher axial compression force than the dorsal instrumentation on the strut graft. Especially during maximal extension there was no axial compression of the dorsal instrumentation on the strut graft, which resulted in an increase of the range of motion. Conclusion. The feasibility of the new method was demonstrated in this study. For the purpose of standardization and comparison it should be considered in spinal implant testing.
  • Keywords
    Spinal stabilization , Spinal decompression , Thoracolumbar burst fractures
  • Journal title
    Clinical Biomechanics
  • Serial Year
    2003
  • Journal title
    Clinical Biomechanics
  • Record number

    486183