Title of article :
Analysis of biomechanical effect of stem-end design in revision TKA using Digital Korean model
Author/Authors :
Yoon Hyuk Kim، نويسنده , , Oh Soo Kwon، نويسنده , , Kyungsoo Kim، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Pages :
6
From page :
853
To page :
858
Abstract :
Background The stem-end pain after revision total knee arthroplasty has been clinically reported around the diaphyseal region of the tibia, and the stem-end design has been recognized as one of the causes of the stem-end pain. But there are few reports about the biomechanical characteristics of different stem-end designs on revision total knee arthroplasty. The objective of this study is to investigate the biomechanical effect of the stem-end design in revision total knee arthroplasty. Methods Three-dimensional finite element models of a tibia and implants for revision total knee arthroplasty with various stem-end design parameters were developed. The tibia model was chosen from 50 male cadavers to represent the standard Korean male. The contact pressures and von Mises stresses around the stem-end were calculated under axial loading. Findings The longer stem length, the bigger stem diameter, the stronger press-fit, and no slot shape increased both peak contact pressures and von Mises stresses. The location of peak contact pressures coincided with the location of stem-end pain reported in clinical studies. Interpretation Different stem-end designs altered the contact pressures or stress concentrations to the tip of the stem. The location of peak contact pressures or von Mises stresses suggested there may be a biomechanical factor to transfer load or generate stress concentration at the location of the stem-end pain in the diaphysis.
Keywords :
Revision total knee arthroplasty , Stem design , Finite element model , Stem-end pain
Journal title :
Clinical Biomechanics
Serial Year :
2008
Journal title :
Clinical Biomechanics
Record number :
486966
Link To Document :
بازگشت