• Title of article

    A new testing rig for force platform calibration and accuracy tests

  • Author/Authors

    H. S. Gill، نويسنده , , J. J. OʹConnor، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1997
  • Pages
    5
  • From page
    228
  • To page
    232
  • Abstract
    Accurate estimation of resultant joint forces and moments from gait analysis data depends heavily on the accuracy of measurement of the ground reaction force (GRF). Typically, multicomponent force platforms are used to measure the components of the GRF and the position of the centre of pressure. A new rig, designed to allow safe and quick static testing of the vertical component and centre of pressure outputs of such force platforms, is described. This rig has been used to test the accuracy of AMTI force platforms and was found to reduce dramatically the testing time. The results of the tests are also given.
  • Keywords
    force platform , Ground reaction force , Gait , Testing
  • Journal title
    Gait and Posture
  • Serial Year
    1997
  • Journal title
    Gait and Posture
  • Record number

    487491