Title of article
Simple gait assessment system
Author/Authors
T. Kar nik، نويسنده , , A. Kralj، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1997
Pages
7
From page
193
To page
199
Abstract
Continuous monitoring and frequent measuring of a patientʹs gait is essential for efficient rehabilitation so that corrective training measures can be adopted throughout the rehabilitation process. Important parameters in gait evaluation are the instantaneous horizontal velocity of the center of gravity (COG) and the overall distance covered. We describe a simple gait assessment system based on combined ultrasound (US) and infra red light (IR) distance/velocity measuring device. The device achieves 0.5% static accuracy within a 10 m range at 25 Hz sampling rate. The system functionality and feasibility was verified in two types of walking: normal gait and functional electrical stimulation (FES) assisted paraplegic gait.
Keywords
rehabilitation , Gait evaluation , Distance and velocity measurement , ultrasound
Journal title
Gait and Posture
Serial Year
1997
Journal title
Gait and Posture
Record number
487517
Link To Document