• Title of article

    Flow injection analytical system for glucose with screen-printed enzyme biosensor incorporating Oscomplex mediator

  • Author/Authors

    Gao، Qiang نويسنده , , Zhang، Chengxiao نويسنده , , Aizawa، Masuo نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2001
  • Pages
    -32
  • From page
    33
  • To page
    0
  • Abstract
    A novel flow injection analytical system with a screen-printed enzyme sensor incorporating Os-complex mediator has been developed. A newly created glucose biosensor was fabricated using thick-film technique and co-immobilization of glucose oxidase and Oscomplex on a permeable membrane. The composition of the enzyme and mediator in the membrane and the analytical conditions were optimized. The flow injection electrochemical system showed that glucose was detected linearly in the concentration range from 0.1 to 10mM with a detection limit of 0.03mM by 50µl of sample injection, giving a throughout of about 40 samples per hour. The glucose biosensors retained their constant response after more than 100 injections and storage over a month. The designed electrochemical flow system with a disposable biosensor is suitable for automatic and rapid determination of glucose.
  • Keywords
    Ion selective electrode (ISE) , Expanded porphyrin , Sapphyrin , Rubyrin , SHG , Apparent surface protonation constant , FTIR-ATR spectrometry , 3,5-Dinitrobenzoate
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2001
  • Journal title
    Analytica Chimica Acta
  • Record number

    49061