Title of article
Mandibular lengthening with an implanted motor-driven device: preliminary study in sheep
Author/Authors
O. Ploder، نويسنده , , Ernst W. Mayr، نويسنده , , G. Schnetz، نويسنده , , E. Unger، نويسنده , , R. Ewers، نويسنده , , H. Plenk Jr، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1999
Pages
4
From page
273
To page
276
Abstract
Distraction osteogenesis with external or intraoral devices is an established method for lengthening human mandibles. In this preliminary study in three sheep a newly designed, fully implantable electromechanical device for mandibular lengthening was used. After osteotomy, the device was screw fixed to the mandible, and the power and control unit was inserted subcutaneously in the neck region. After a healing period of five days, the device was activated magnetically, allowing calibrated distraction steps of 0.04 mm/h, resulting in a total of 1.0 mm/day. Over a period of 14 days, a maximum mandibular lengthening of 13.6 mm could be achieved without transmucosal activation. Depending on stability of the screw fixation, membranous and/or cartilaginous bone formation was observed in the callus by radiological and histological evaluation. Further experimental research is ongoing to prove the clinical usefulness of this device.
Journal title
British Journal of Oral and Maxillofacial surgery
Serial Year
1999
Journal title
British Journal of Oral and Maxillofacial surgery
Record number
492545
Link To Document