Comparison of two systems for rigidly connecting 2.0-mm bone screws to an implantable device: in vitro stability testing
Author/Authors
J. -P. van Loon، نويسنده , , L. G. M. de Bont، نويسنده , , G. J. Verkerke، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2000
Pages
5
From page
200
To page
204
Abstract
The stability of a screw-fixed implantable device can be improved by eliminating the freedom of movement between the screws and the device. Two systems have been developed for rigidly connecting 2.0-mm bone screws to an implantable device, and the aim of this study was to test and compare the stability of the two systems.
In system A, a fixing disc locked the screw-head. In system B, the thread of the screw locked directly into the device. The stability of the connection was tested by measuring the resistance against a torque on the screw, against shear loading, against push-out loading, and against dynamic shear loading (70 N) for 5 million cycles.
Both systems met the required minimum values for the resistance against a torque, shear loading and push-out loading, and dynamic shear loading did not cause movements between the screw and the device.
We conclude that both systems are stable, but prefer system B because it is easier to make and implant than system A.