• Title of article

    Flow-injection spectrophotometric determination of cobalt(II) at low µg l-1 levels with 4benzylpiperidinedithiocarbamate

  • Author/Authors

    Andac، M. نويسنده , , Asan، A. نويسنده , , Isildak، I. نويسنده , , Cesur، H. نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2001
  • Pages
    -142
  • From page
    143
  • To page
    0
  • Abstract
    A rapid, sensitive and reproducible flow-injection (FI) analysis method is described for the spectrophotometric determination of cobalt(II) using 4-benzylpiperidinedithiocarbamate as a chromogenic reagent. The method involves the measurement of the absorbance of the colored cobalt(II)-(4-benzylpiperidinedithiocarbamate)2 complex at 640nm. In the system, a microcolumn (MC) containing strong cation-exchange resin was placed between injection manifold and spectrophotometer to increase interactions between cobalt(II) and reagent, and for preconcentration of cobalt(II). The linear range of the proposed FI method was 0.6-100 µgl-1, and the limit of detection (signal/noise=3) was 0.1µgl-1 for a 10µl injection of cobalt(II) ion solution at a sampling rate of 60h-1. The relative standard deviation (R.S.D.) for 0.6-100 µgl-1 cobalt(II) was <1.0% (n=5). The precision and accuracy of the method were checked by analysis of certified reference metals, and the method was applied to river, sea and hot spring water samples.
  • Keywords
    Beverages analysis. , Potentiometry , Benzoate sensitive electrode
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2001
  • Journal title
    Analytica Chimica Acta
  • Record number

    49329