Title of article :
Electroanalytical determination of acid phosphatase activity by monitoring p-nitrophenol
Author/Authors :
Aoyama، Hiroshi نويسنده , , Kubota، Lauro T. نويسنده , , Calvo-Marzal، Percy نويسنده , , Rosatto، Simone S. نويسنده , , Granjeiro، Paulo A. نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Abstract :
The development of a modified electrode for the selective determination of p-nitrophenol (PNP) to be applied in the determination of the enzymatic activity (EA) of acid phosphatase using p-nitrophenylphosphate (PNPP) as substrate is described. Elimination of the interference of PNPP was made by covering the electrode surface using 25µl of a 5% (m/v) Nafion® solution. In comparison to the glassy carbon (GC) electrode without modification, the presence of the membrane promoted a reduction in the sensitivity and an increase in the detection limit for the determination of PNP, in addition to a decrease of the electron transfer constant. The GC electrode covered with a Nafion® membrane showed a linear response range between 20 and 230µmoll-1 for PNP, adjusted by the equation IP=0.157+0.0089 [PNP], for N=22 with a correlation coefficient of 0.9984. This electrode was used for the determination of the EA of acid phosphatase from castor bean seeds and an activity of 38.40Umg-1 was found
Keywords :
urea , Pyruvate oxidase , Cathodic detection
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta