Load distribution and the predictive power of morphological indices in the distal radius and tibia by high resolution peripheral quantitative computed tomography
Author/Authors :
Joshua A. MacNeil، نويسنده , , Steven K. Boyd، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Pages :
9
From page :
129
To page :
137
Abstract :
A 3D high resolution peripheral quantitative computed tomography scanner (HR-pQCT) (XtremeCT, Scanco Medical, voxel size 82 μm) has been recently developed that can perform in vivo human measurements on peripheral sites, including the wrist and tibia. The goals of this study were to use HR-pQCT measurements to determine the ability of morphological and density measurements to predict bone apparent stiffness and apparent Youngʹs modulus in the distal radius and tibia, to determine the relative importance of cortical and trabecular bone in carrying load in the human distal radius and tibia. Furthermore, the ability of a sub-volume of trabecular bone apparent Youngʹs modulus to predict the Youngʹs modulus of a whole radius and tibia section was determined. A total of 25 measurements of the radius and 12 measurements of the tibia were used for morphological and finite element analyses of sections, and sub-volume cubes of trabecular bone from the distal radius and tibia. The subjects were chosen to obtain a large variation in age ranges and bone architecture and density. By combining multiple measurements, a strong ability to predict bone apparent stiffness and apparent Youngʹs modulus was found for morphological and density measurements in the radius and tibia (R2 > 0.80). The relative importance of the trabecular and cortical bone in carrying load was also found to vary consistently with location in the sample for both the radius and the tibia. This indicates that measurements of the cortical and trabecular bone are required for assessing fracture risk. A cubic section of trabecular bone was found to be insufficient to accurately represent the apparent bone Youngʹs modulus of a radius or tibia section. Morphological and density measurements of the distal radius and tibia have been shown in this study to predict bone apparent Youngʹs modulus and apparent stiffness, and may indicate when a more time consuming finite element analysis is warranted. It should be noted that these results may be an overestimation of the predictive ability of structural parameters, as the influence of bone density is removed from the finite element analyses, and the results were only influenced by bone structure. A measurement of bone apparent Youngʹs modulus is independent of subject size (as opposed to reaction force), and may provide the ability to distinguish between two patients that have similar mean morphological and density measurements; but different overall structures, and therefore, different fracture risk.
Keywords :
Bone quality , Finite element analyses , Peripheral quantitative computed tomography , Load distribution , Osteoporosis
Journal title :
Bone
Serial Year :
2007
Journal title :
Bone
Record number :
496456
Link To Document :
بازگشت