Title of article :
Simultaneous detection of cadmium, copper, and lead using a carbon paste electrode modified with carbamoylphosphonic acid self-assembled monolayer on mesoporous silica (SAMMS)
Author/Authors :
Yantasee، Wassana نويسنده , , Lin، Yuehe نويسنده , , Fryxell، Glen E. نويسنده , , Busche، Brad J. نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Abstract :
A new sensor was developed for simultaneous detection of cadmium (Cd2+), copper (Cu2+), and lead (Pb2+), based on the voltammetric response at a carbon paste electrode modified with carbamoylphosphonic acid (acetamide phosphonic acid) self-assembled monolayer (SAM) on mesoporous silica (Ac-Phos SAMMS). The adsorptive stripping voltammetry (AdSV) technique involves preconcentration of the metal ions onto Ac-Phos SAMMS under an open circuit, then electrolysis of the preconcentrated species, followed by a square wave potential sweep towards positive values. Factors affecting the preconcentration process were investigated. The voltammetric responses increased linearly with the preconcentration time from 1 to 30 min or with metal ion concentrations ranging from 10 to 200 ppb. The responses also evolved in the same fashion as adsorption isotherm in the pH range of 2–6. The metal detection limits were 10 ppb after 2 min preconcentration and improved to 0.5 ppb after 20 min preconcentration.
Keywords :
Adsorptive stripping voltammetry , Chemically modified electrode , Self-assembled monolayer , carbon paste electrode , Carbamoylphosphonic acid
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta