Title of article :
Recalibration of a gas-sensor array system related to sensor replacement
Author/Authors :
Eklov، Tomas نويسنده , , Tomic، Oliver نويسنده , , Kvaal، Knut نويسنده , , Haugen، John-Erik نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Pages :
-198
From page :
199
To page :
0
Abstract :
Univariate multiplicative drift correction and multivariate component correction were applied for recalibration of long-term measurement data acquired with a solidstate gas-sensor array system. The efficiency of the methods was evaluated by classifying recalibrated measurement data using k-nearest neighbor classification and partial least-squares discriminant analysis. For the measurement data in this experiment both multiplicative drift correction and component correction appeared to be useful for recalibration of measurement data from the new gas-sensor array with regard to measurement data acquired with the old replaced gas-sensor array.
Keywords :
Multiplicative drift correction , Component correction , Gas-sensor array , Electronic nose , Recalibration , Sensor replacement