Title of article :
Optimization of a nitrogen analyser based on the Dumas method
Author/Authors :
Saint-Denis، Thierry نويسنده , , Goupy، Jacques نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Pages :
-190
From page :
191
To page :
0
Abstract :
The objectives of this work was to obtain a total nitrogen concentration in milk with the Dumas method for industrial and research applications. This method was faster than the Kjheldal method (5 min against 2 h or 3 h) but less precise. A three factor experimental design was performed to optimize the instrument, a FP-2000 supplied by LECO. The first interpretation of the experimental design was disappointing and no conclusion could be done. It was not a failure of experimental design but a lack of thinking on the physical and chemical aspect of the problem. Selection and construction of composite responses based on chemical and physical consideration were the keys to optimize the Dumas method. This method is now as precise as the Kjheldal method, but considerably faster.
Keywords :
Nitrogen , Dumas , Experimental designs
Journal title :
Analytica Chimica Acta
Serial Year :
2004
Journal title :
Analytica Chimica Acta
Record number :
50372
Link To Document :
بازگشت