Title of article :
Variation of depth of cure and intensity with distance using LED curing lights
Author/Authors :
Krishna Aravamudhan، نويسنده , , Duane Rakowski، نويسنده , , P.L. Fan، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Abstract :
Objectives
The purpose of the study was to determine the correlation between intensity of light-emitting diode (LED) and tungsten-halogen light sources, and depth of cure of a resin composite at different distances.
Methods
Four LED curing lights (Flashlite 1001, Freelight 2, Smartlite IQ and Ultralume 5) and one tungsten halogen (Optilux 501, with 8 and 11 mm tips) were evaluated. Intensity was measured according a modified ISO Standard 10650 at distances of 0, 2, 4, 6, 8, 10 mm between the light tip and detector. Depth of cure (DOC) of TPH Spectrum shade A2 was measured according to the international standard ISO 4049 at the same distances.
Results
For all lights, intensity decreased as distance increased. The authors documented a logarithmic correlation between intensity and distance for all lights except the Smartlite IQ, Ultralume 5 and the Optilux 501 with the 11 mm tip, which showed a linear relationship between intensity and distance. All lights demonstrated a logarithmic correlation between intensity and DOC, and a linear correlation between DOC and distance. Smartlite IQ and Optilux 501 (11 mm tip) also had the least reduction in intensity and DOC at 10 mm.
Significance
Clinicians often an experience difficulty placing the light tip close to the resin surface when curing resin composites. While both intensity and DOC decrease with increasing distance, the relationship between these factors and distance may not be similar for all lights and may depend on the characteristics of individual lights.
Keywords :
Intensity , Depth of cure , Distance , Dental material , Resin Composite , Light emitting diode , Light curing