    Microfabricated On-Chip-Type Electrochemical Flow Immunoassay System for the Detection of Histamine Released in Whole Blood Samples

  • Author/Authors

    Lim، Tae-Kyu نويسنده , , Matsunaga، Tadashi نويسنده , , Ohta، Hiroko نويسنده ,

  • Issue Information
    دوهفته نامه با شماره پیاپی سال 2003
  • Pages
    -3315
  • From page
    3316
  • To page
    0
  • Abstract
    This paper describes an on-chip-type electrochemical flow immunoassay system with a multichanneled matrix column. The multichanneled matrix column was functionally coated with cation-exchange resin and used for separation of proteins. Antihistamine immunoglobulin G (IgG) antibody conjugated with ferrocenemonocarboxylic acid (Fc) was also prepared and used as a novel analytical reagent. Antibody-antigen complexes were separated from free Fc-conjugated IgG antibody (Fc-IgG) on the basis of differences in isoelectric point (pI) using the multichanneled matrix column coated with cation-exchange resin. The assay yields a good relationship between current and histamine concentration in the range of 200-2000 ng/mL. This simple technique enables the assay of histamine released in whole blood within 2 min. Furthermore, a good correlation was found between the response of the electrochemical immunoassay described in this paper and the conventional RIA (radioimmunoassay). This on-chip-type electrochemical flow immunoassay requires only minute quantities of whole blood samples and generates highly reproducible results.
  • Keywords
    Crop yields , Yield gains , Hedges , Shelterbelts , Field margins
  • Journal title
    Analytical Chemistry
  • Serial Year
    2003
  • Journal title
    Analytical Chemistry
  • Record number