Title of article :
Direct Determination of Carbohydrates, Amino Acids, and Antibiotics by Microchip Electrophoresis with Pulsed Amperometric Detection
Author/Authors :
Garcia، Carlos D. نويسنده , , Henry، Charles S. نويسنده ,
Issue Information :
دوهفته نامه با شماره پیاپی سال 2003
Abstract :
The separation and detection of underivatized carbohydrates, amino acids, and sulfur-containing antibiotics in an electrophoretic microchip with pulsed amperometric detection (PAD) is described. This report also describes the development of a new chip configuration for microchip electrophoresis with PAD. The configuration consists of a layer of poly(dimethylsiloxane) that contains the microfluidic channels, reservoirs, and a gold microwire, sealed to a second layer of poly(dimethylsiloxane). Example sepa-rations of carbohydrates, amino acids, and sulfur-containing antibiotics are shown. The effect of the separation and injection potentials, buffer pH and composition, injection time, and PAD parameters were studied in an effort to optimize separations and detection. Detection limits ranging from 6 fmol (5 (mu)M) for penicillin and ampicillin to 455 fmol (350 (mu)M) for histidine were obtained.
Keywords :
Shelterbelts , Hedges , Field margins , Yield gains , Crop yields
Journal title :
Analytical Chemistry
Journal title :
Analytical Chemistry