• Title of article

    Fluidic Preconcentrator Device for Capillary Electrophoresis of Proteins

  • Author/Authors

    Astorga-Wells، Juan نويسنده , , Swerdlow، Harold نويسنده ,

  • Issue Information
    دوهفته نامه با شماره پیاپی سال 2003
  • Pages
    -5206
  • From page
    5207
  • To page
    0
  • Abstract
    A new preconcentration device was developed for analysis of proteins by capillary electrophoresis (CE). The microfluidic device uses an electric field to capture proteins that pass through the system. The capture zone is maintained in the flow stream by the interaction between hydrodynamic and electrical forces. The device consists of a flow channel made of PEEK tubing with two electrical junctions, each of which is covered with a conductive membrane. A syringe pump provides the flow stream and also allows the injection of up to 13.5 (mu)L of a dilute sample. The system can be easily connected to a CE device post-capture for off-line preconcentration of proteins. For the proteins used in this study, preconcentration factors up to 40-fold can be achieved. CE detection limits for bovine carbonic anhydrase, (alpha)-lactalbumin and (beta)-lactoglobulins A and B were in the nanomolar range using UV detection at 200 nm. Preconcentration is dependent on both time and initial protein concentration. We show the possibility of using an off-line fluidic preconcentrator device employing counterflow capillary electrophoresis with minimum sample manipulation, achieving detection limits similar to on-line approaches.
  • Keywords
    gas_phase measurement , particle_phase measurement
  • Journal title
    Analytical Chemistry
  • Serial Year
    2003
  • Journal title
    Analytical Chemistry
  • Record number