Title of article :
Comparison of standard and pediatric size concentric needle EMG electrodes
Author/Authors :
Alexander A. Brownell، نويسنده , , Mark B. Bromberg، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Pages :
4
From page :
1162
To page :
1165
Abstract :
Objective To compare motor unit action potential (MUAP) metrics recorded by standard and pediatric size concentric EMG electrodes. Methods Commercial electrodes were used to record MUAPs from biceps brachii, first dorsal interosseous and tibialis anterior muscles in normal subjects and those with amyotrophic lateral sclerosis (ALS). Results In normal subjects, peak amplitude and area were significantly higher when recorded by the pediatric size electrode in tibialis anterior muscles and peak amplitude recorded in first dorsal interosseous muscles. In ALS subjects, peak amplitude was higher recorded by the pediatric size electrode in tibialis muscle but lower when recorded in first dorsal interosseous muscles. Conclusions Differences of MUAP metrics when recording with standard and pediatric size electrodes do not seem to have a clinical relevance. Significance Pediatric and standard concentric electrodes record similar MUAP metrics
Keywords :
EMG , Concentric electrodes , Motor unit action potential , MUAP , Electrode comparison
Journal title :
Clinical Neurophysiology
Serial Year :
2007
Journal title :
Clinical Neurophysiology
Record number :
523955
Link To Document :
بازگشت