Author/Authors :
Emile G. Daoud MD، نويسنده , , William Kou، نويسنده , , Theresa Davidson، نويسنده , , Mark Niebauer، نويسنده , , Frank Bogun، نويسنده , , Mark Castellani، نويسنده , , Keung K. Chan، نويسنده , , Rajiva Goyal، نويسنده , , Mark Harvey، نويسنده , , S. Adam Strickberger، نويسنده , , K. Ching Man، نويسنده , , Fred Morady، نويسنده ,
Abstract :
The Accufix (Telectronics Pacing Systems, Englewood, Colo.) atrial active fixation pacemaker lead has a preformed J stiff retention wire than can fracture and erode through the atrium. We screened 85 patients with digital fluoroscopy and detected 18 (21%) lead fractures. Fourteen (16%) leads were successfully extracted by percutaneous techniques. The mean time for extraction and placement of a new atrial lead was 109 ± 53 min, and a lead> 1 year old but not in the presence of a fracture was associated with a significantly longer extraction time. There were two extraction complications. One patient had a fever following the extraction, and one patient had a focal sensory deficit 1 week after the extraction probably caused by a paradoxical embolus. Standard percutaneous extraction tools can be used for successful and safe removal of a fractured Accufix lead.