• Title of article

    Safety and feasibility of continuous monitoring of coronary flow velocity in acute myocardial infarction by Doppler guide wire

  • Author/Authors

    Masato Nakamura، نويسنده , , Taro Tsunoda، نويسنده , , Tetsuzo Wakatsuki، نويسنده , , Katsuto Ui، نويسنده , , Toshiyuki Degawa، نويسنده , , So Yabuki، نويسنده , , Tetsu Yamaguchi، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1996
  • Pages
    5
  • From page
    567
  • To page
    571
  • Abstract
    We investigated the safety and limitations of the Doppler guide wire for continuous monitoring of coronary flow velocity outside the catheterization laboratory in 17 patients with acute myocardial infarction. After direct angioplasty, patients were taken to the coronary care unit with the Doppler guide wire positioned within the artery. Flow velocity was continuously monitored. Heparin was titrated to an active clotting time>200 seconds. Clinical outcome and angiographic analyses were evaluated. Flow velocity monitoring was conducted with an 88.2% success rate and lasted for 16 ± 5 hours. Monitoring failed in 2 of the 4 right coronary artery cases. Small amounts of thrombus were seen to adhere to the Doppler guide wire at the end of monitoring. No complications were related to the procedure. No deterioration of angiographic findings was observed. This preliminary study confirmed the safety of the Doppler guide wire for continuous monitoring of coronary flow in patients with acute myocardial infarction.
  • Journal title
    American Heart Journal
  • Serial Year
    1996
  • Journal title
    American Heart Journal
  • Record number

    527097