Title of article :
Overdrive atrial pacing for conversion of atrial flutter: Comparison of postoperative with nonpostoperative patients
Author/Authors :
Robert W. Peters، نويسنده , , Steven R. Shorofsky، نويسنده , , Michael Pelini، نويسنده , , Mary Olsovsky، نويسنده , , Michael R. Gold، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Pages :
4
From page :
100
To page :
103
Abstract :
Background Previous studies have reported varying success rates in overdrive pace termination of atrial flutter. We hypothesized that these discrepancies might be caused by differences in study populations. Accordingly, we prospectively compared the success rate of pacing in patients with atrial flutter that occurred after heart surgery with that of patients with atrial flutter from other causes. Methods and Results The study population consisted of 65 consecutive patients referred for pace termination of typical (type I) atrial flutter. Pacing was performed in 30-second bursts, starting at the flutter cycle length, and repeated in 5-ms decrements until normal sinus rhythm or atrial fibrillation occurred. Normal sinus rhythm was restored in 38 (65%) patients. Of 20 patients whose flutter was precipitated by heart surgery, 19 (95%) were successfully pace terminated. In contrast, pace termination was successful in only 47% of the remainder of the population (P < .001). No other clinical parameters were predictive of outcome. Conclusions We conclude that overdrive pacing is an effective means of terminating atrial flutter that has occurred after heart surgery. Alternative methods should be considered as the initial therapeutic approach in patients with atrial flutter from other causes. (Am Heart J 1999;137:100-3.)
Journal title :
American Heart Journal
Serial Year :
1999
Journal title :
American Heart Journal
Record number :
531441
Link To Document :
بازگشت