Title of article
Dose intercomparison experiment for gamma rays and 3-MeV electrons by mailing dosimetry using free-radical dosimeters
Author/Authors
S. R. Nilekani، نويسنده , , G. R. Narayan، نويسنده , , B. Suseela، نويسنده , , R. M. Bhat، نويسنده , , R. B. L. Gupta and A. Shivpuri، نويسنده , , Takuji Kojima، نويسنده , , Haruki Takizawa، نويسنده , , Hiromi Sunaga، نويسنده , , Ryuichi Tanaka، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1995
Pages
3
From page
205
To page
207
Abstract
The high-dose intercomparison experiments between BARC and JAERI for 60Co-γ-rays and 3-MeV electrons were carried out by mailing alanine (powder and alanine-polymer) and glutamine (powder) dosimeters between the two facilities. Powder dosimeters and alanine-polymer dosimeters were read out by spectrophotometry and EPR spectrometry, respectively. Absorbed dose values at about 1, 10, 33 and 54 kGy estimated by these dosimeters were in good agreement with JAERI nominal doses within ±5% both for γ-rays and electrons. This preliminary result shows that these free-radical dosimeters have useful characteristics as mailing dosimeters for regional transfer dose standardization.
Journal title
Applied Radiation and Isotopes
Serial Year
1995
Journal title
Applied Radiation and Isotopes
Record number
539200
Link To Document